For a sales quote, meeting request, evaluation unit, or other sales question, please fill in the form below:
|
Single-Output Precision High Voltage Power Supply Module
Typical Applications |
Features |
-
Bias supplies
-
Mass spectrometry
-
SEM/FIB
-
Electron beams
-
Ion beams
|
-
Precision output voltage from
-
0 to 1 kV through 0 to 15 kV
-
PPM level ripple, regulation and stability
-
As low as 10 ppm temperature coefficient and reference
-
0 to 4, 15/20, or 30 W of output power
-
Maximum Iout capability down to 0 V
-
Voltage and current regulation/limit capability
-
Precision output voltage and current monitors
|
Specifications
Input voltage range |
23 VDC to 30 VDC full power |
Input current |
Standby/disable: <50
No load, max EOUT: <325
Full load, max EOUT: 2.5 |
AC ripple current |
<10 nominal input, full load |
Output |
1E |
2E |
4E |
6E |
10E |
15E |
Output voltage range |
0 to 1,000
VDC |
0 to 2,000
VDC |
0 to 4,000
VDC |
0 to 6,000
VDC |
0 to 10,000
VDC |
0 to 15,000
VDC |
Nominal input voltage/model |
24 |
24 |
24 |
24 |
24 |
24 |
Output power (W) |
4 |
20 |
30 |
4 |
20 |
30 |
4 |
20 |
30 |
4 |
20 |
30 |
4 |
15 |
30 |
4 |
15 |
30 |
Output current (mA) |
4 |
20 |
30 |
2 |
10 |
15 |
1 |
5 |
7.5 |
0.67 |
3.3 |
5 |
0.4 |
1.5 |
3 |
0.26 |
1 |
2 |
Voltage monitor |
0 to 10 ±0.5% for normal operating conditions |
Current monitor |
0 to 10 ±0.5% for normal operating conditions |
Ripple |
≤10 for full load, max EOUT |
Line regulation |
<25 ppm or <10 ppm for nom input, max EOUT, full power |
Static load regulation |
<25 ppm or <10 ppm for no load to full load, max EOUT |
Stability |
<25 ppm or <10 ppm for 30 min warmup, per 8 h, per day |
Programming and Controls
Input impedance |
10 for nominal input |
Adjust accuracy and adjust linearity |
±0.05% for 10 to 100% |
Adjust voltage |
0 to +10 for differential |
Output voltage |
+10.00 ±0.05% for T = +25°C, Initial Value |
Max source current |
5 for T = +25°C |
Output impedance |
Buffered, low impedance, 2 mA max for source/sink current for normal operating conditions |
Enable/disable |
0 to +0.8 disable, +2.5 to 10 enable (default = disable) |
Environmental
Operating |
+10 to +45 for full load, max EOUT, case temp. |
Temperature coefficient |
±25 or ±10 over the specified temperature |
Thermal shock |
-40 to +65 for Mil-Std-810, Method 504, Class 2 |
Storage |
-55 to +105 for non-operating, case temp. |
Humidity |
0 to 95%, non-condensing for all conditions, standard package |
Altitude |
Sea level through 10,000 for standard package, all conditions |
Shock |
20 for Mil-Std-810, Method 516, Proc. 4 |
Vibration |
10 for Mil-Std-810, Method 514, Fig. 514-3 |
Accuracy of specifications in unlocked documents cannot be guaranteed. Please contact your AE representative to confirm specifications.
|