For a sales quote, meeting request, evaluation unit, or other sales question, please fill in the form below:
|
Mass Spectrometry Power Suplies
The MSRD series consists of high-stability, reversible source modules that perform reliably even under short-circuit or arc conditions. It pairs fast, reversible bias output with a floating +3 kV, 400 μA detector output, offers a differential programming input, and can be set by an external voltage.
The modular design of AE high voltage products for mass spectrometry enables an array of performance features and combinations. From simple options, such as cable length and connector type, to complete custom designs, we deliver solutions that precisely fulfill your specific requirements.
Features |
-
Output power: 1.8 W, reversible
-
Output voltage: 0 to ±12 kV (bias), 0 to +3 kV (detector)
-
Ripple: < 200 mV (bias and detector)
-
Temperature coefficient: 25 ppm per°C (bias), 300 ppm per °C (detector); lower values available
-
Stability: 100 ppm (bias)
-
Variable floating detector: 1.2 W
-
Bias four quadrant output stage
-
Bias controllable through zero
-
High reliability
|
Specifications
Output Power |
Bias: 1.8 W max
Detector: 1.2 W max |
Output Voltage |
Bias: 0 to ±12 kV
Detector: 0 to +3 kV |
Output Current |
Bias: 150 to 250 μA, depending on model
Detector: 400 μA |
Input Coltage |
+24 VDC ±10% |
Input Current |
1.4 A max, depending on model |
Line Regulation |
Bias: < 1.2 V
Detector: < 3 V for a 5% input voltage change |
Load Regulation |
Bias: < 1.2 V
Detector: < 3 V for a 10 to 100% load change |
Ripple |
Bias: < 200 mV
Detector: < 200 mV |
Voltage Control |
Bias: 0 to +10 V for full scale accuracy ±2%
Detector: 0 to +10 V for full scale accuracy ±5% |
Current Control |
Bias: fixed at 100 μA above nominal
Detector: fixed at 500 μA |
Voltage Monitor |
Bias: 0 to ±10 V for full scale accuracy ±2%
Detector: 0 to +10 V for full scale accuracy ±5 % |
Current Monitor |
Bias: 0 to ±10 V for full scale accuracy ±5%
Detector: -- |
Polarity Control |
Bias output, < 1 V = negative output, > 3.5 V or open = positive output |
Inhibit |
Bias and detector outputs < 1 V or open = off/inhibited, > 3.5 V = operating |
Stability |
Bias: 100 ppm
Detector: 1000 ppm over a 1 hr period |
Limits |
Units have limits for over-voltage, short-circuit, and intermittent arcs to ground. |
Reliability |
MTBF > 50,000 hr |
Humidity |
80% max relative humidity up to 31°C (88°F), reducing linearly to 50% at 40°C (104°F); non-condensing (ref EN61010-1) |
Safety |
Meets the requirements of the low-voltage directive, 2006/95/EC by complying with BS EN61010-1:2010 when installed as a component part of compliant equipment. Units are CE-marked accordingly. |
RoHS |
Meets the requirements of EU directive 2011/65/EC on the restriction of use of certain hazardous
substances in electrical and electronic equipment (RoHS) |
Construction |
A fabricated aluminium alloy case is used for good heat dissipation and screening. |
Environmental
Temperature Coefficient |
Bias: 25 ppm per °C (lower values available upon request)
Detector: 300 ppm per °C (lower values available upon request) |
Cooling |
Convection-cooled |
Operating Temperature |
10 to 50°C (50 to 122°F) |
Storage/Transport Temperature |
-20 to 85°C (-4 to 185°F) |
Operational Altitude |
Sea level to 2000 m (6500′) |
Storage/Transport Altitude |
Sea level to 18,000 m (59,055′) |
Accuracy of specifications in unlocked documents cannot be guaranteed. Please contact your AE representative to confirm specifications.
This series is available in the following models.
Model |
Input
Voltage |
Output
Voltage |
Output
Current (Max) |
Output
Power |
MSRD-502 |
+24 VDC |
±5 kV |
250 µA |
1.8 W |
MSRD-752 |
+24 VDC |
±7.5 kV |
200 µA |
1.8 W |
MSRD-103 |
+24 VDC |
±10 kV |
175 µA |
1.8 W |
MSRD-123 |
+24 VDC |
±12 kV |
150 µA |
1.8 W |
|